Description
Over the past few years, with the popularity of cooking shows, there has been an increasing trend of people experimenting more with flavors and tastes. One way that people experiment in the kitchen is through the use of spices and herbs. Considering the International Organization for Standardization (ISO) identifies over 100 varieties of herbs and spices, there are many different options for flavor combinations.
Spices and herbs have been traded throughout the world for use in food seasoning and coloring for thousands of years. Traditionally, spices were processed and dried by being laid in the sun to dry. Since then, much more stringent standards have been established. These standards ensure that the final product is unadulterated and meets health and safety standards. These standards are set by several governing bodies, such as the American Spice Trade Association (ASTA) and the European Spice Association (ESA). According to both of these regulatory agencies, moisture content is an important parameter to measure. Monitoring moisture content in spices helps to prevent mold, bacteria, and yeast growth. Moisture content can also be monitored over time to assess storage conditions. The packaging in which the spices are stored will directly affect moisture uptake. For example, a package made out of paper might be sufficient in a dry environment, while spices in a humid environment may require additional protection. Based on the moisture content of the spices, adjustments to packaging materials or the environment in which spices are manufactured can be made as needed.
Application
A global spice importer, processor, and purveyor
contacted Hanna about aiding in their quality
control process. This customer had been
sending samples out for moisture analysis and
wanted to bring the analysis into their onsite
quality assurance lab. Hanna helped the
customer to develop an extraction method for
a variety of spices that would be run on the
HI903 Volumetric Karl Fischer Titrator. It
was determined that an extraction in methanol
would effectively release the moisture in all of
the spices being analyzed. Due to the variety
of spices being run on the HI903, an overnight
extraction was suggested to ensure all spices
had ample time release their water; this
allowed the same procedure to be used for all
samples, which increased the confidence that all
extractions would be carried out correctly.
The customer appreciated that the external
extraction parameters could be entered
directly into the titrator and the calculation
is automatically performed. This saved them valuable time and reduced any errors from
performing a manual calculation. The USB port
of the HI903 allowed for easy transfer of the
results by using a common flash drive. The
customer also appreciated the sealed solvent
exchange system that allows the solvent in the
titration cell to be exchanged without opening the cell. This system uses an air pump to add
and remove solvent without exposing the
cell to atmospheric moisture. With the help of
Hanna Instruments’ Applications Team and the
features of the HI903 Karl Fischer, this customer
was able to bring moisture analysis into their
QA lab.
Comments